Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

CNA Affected

[
  {
    "product": "Omnipod Insulin Management System",
    "vendor": "Insulet",
    "versions": [
      {
        "status": "affected",
        "version": "19191"
      },
      {
        "status": "affected",
        "version": "40160"
      },
      {
        "status": "affected",
        "version": "ZXP425"
      },
      {
        "status": "affected",
        "version": "ZXR425"
      }
    ]
  }
]

7.3 High

CVSS3

Attack Vector

LOCAL

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

LOW

Integrity Impact

HIGH

Availability Impact

LOW

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L

8.1 High

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

26.7%

Related for CVELIST:CVE-2020-10627