An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.
7.8CVSS
7.7AI Score
0.0004EPSS
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access.
6.8CVSS
6.5AI Score
0.001EPSS
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without integrity protection against tampering.
7.5CVSS
7.5AI Score
0.002EPSS
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against tampering.
5.5CVSS
5.6AI Score
0.0004EPSS