CVE-2017-9657
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor,...
6.5CVSS5.5AI score0.00794EPSS
SaveExploits0References4
Unspecified Vulnerability in Philips' IntelliView MX40 Patient Worn Monitor (CNVD-2017-26427)
The MX40 Patient Worn Monitor is primarily used as a traditional telemetry medical device as part of a surveillance and alarm system. Under certain 802.11 network conditions, when hospital staff reset the device and re-establish it to a Wi-Fi access point, the MX40 can connect to an alternate...
6.5CVSS6.4AI score0.00837EPSS
SaveExploits0References1