CVE-2020-12037
CVE-2020-12037 affects Baxter PrismaFlex (all versions) and PrisMax (all versions before 3.x). The root cause is a hard-coded service password (CWE-259), enabling an attacker to access device settings, calibration data, and network configuration. The ICS update also notes lack of data-in-transit ...
Baxter PrismaFlex Hardcoding Vulnerability
The Baxter PrismaFlex is a critical care device from Baxter. A hard-coded vulnerability exists in Baxter PrismaFlex all versions that stems from the fact that PrismaFlex contains a hard-coded service password that can be exploited by an attacker to modify device settings and calibration values...
Baxter PrismaFlex and PrismMax Trust Management Issues Vulnerabilities
The Baxter PrismaFlex and PrismMax are both critical care devices from Baxter, Inc. A trust management issue vulnerability exists in Baxter PrismaFlex all versions and PrismMax prior to version 3.x. The vulnerability stems from a lack of authentication on the device and can be exploited by an...
Baxter PrismaFlex and PrismMax Information Disclosure Vulnerabilities
The Baxter PrismaFlex and PrismMax are both critical care devices from Baxter. An information disclosure vulnerability exists in Baxter PrismaFlex all versions and PrismMax prior to version 3.x. The vulnerability stems from the failure of an affected device to encrypt e.g., TLS/SSL transmitted da...
Baxter PrismaFlex and PrisMax (Update B)
1. EXECUTIVE SUMMARY CVSS v3 7.6 ATTENTION: Exploitable remotely/low skill level to exploit Vendor: Baxter Equipment: PrismaFlex and PrisMax Vulnerabilities: Cleartext Transmission of Sensitive Information, Improper Authentication, Use of Hard-Coded Password 2. UPDATE INFORMATION This updated...