EUVD-2025-10702
Malicious code in bioql PyPI...
The vulnerability of the .NET Remoting technology in systems for visualization and analysis of clinical data based on artificial intelligence, IntelliSpace Portal. This allows a intruder to execute arbitrary code.
The vulnerability of the .NET Remoting technology in systems for visualization and analysis of clinical data based on artificial intelligence, IntelliSpace Portal, is related to deficiencies in the deserialization mechanism. Exploiting this vulnerability could allow an attacker to execute arbitra...
CVE-2025-3425
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the deserialization vulnerability. After analyzing the configuration files, we observed that the server had set the TypeFilterLevel to Full which is...
CVE-2025-3424
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the "Object Marshalling" technique, which allows an attacker to read internal files without any authentication. This is possible by crafting specifi...
CVE-2025-3424
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the "Object Marshalling" technique, which allows an attacker to read internal files without any authentication. This is possible by crafting specifi...
CVE-2025-3425
The CVE-2025-3425 issue affects Philips IntelliSpace Portal (12 and earlier). It arises from a .NET Remoting deserialization vulnerability exposed via port 755, where TypeFilterLevel was configured to Full, potentially enabling remote code execution. Impact is described as remote code execution w...
CVE-2025-3424
The CVE-2025-3424 issue affects Philips IntelliSpace Portal (versions 12 and prior) and stems from the .NET Remoting-based access on port 755 via Object Marshalling, enabling an unauthenticated attacker to read internal files by crafting specific Remoting URLs derived from client-side configurati...
Philips IntelliSpace Portal 路径遍历漏洞
Philips IntelliSpace Portal is an advanced medical image analysis platform from Philips Netherlands that provides multimodal image processing and automated diagnostic tools. A path traversal vulnerability exists in Philips IntelliSpace Portal version 12 and earlier, which stems from the...
PT-2025-15255 · Philips · Intellispace Portal
Name of the Vulnerable Software and Affected Versions: IntelliSpace Portal versions 12 and prior Description: The issue arises from the exploitation of port 755 through the "Object Marshalling" technique, allowing an attacker to read internal files without authentication. This is possible by...
PT-2025-15269 · Philips · Intellispace Portal
Name of the Vulnerable Software and Affected Versions: IntelliSpace Portal versions 12 and prior Description: The issue arises from the exploitation of port 755 through a deserialization vulnerability in the IntelliSpace portal application, which utilizes .NET Remoting for its functionality. The...