Lucene search
+L

9 matches found

Prion
Prion
•added 2019/03/25 5:29 p.m.•30 views

Hardcoded credentials

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

10CVSS6.9AI score0.01972EPSS
SaveExploits0References1Affected Software3
NVD
NVD
•added 2019/03/25 5:29 p.m.•27 views

CVE-2015-3954

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira...

10CVSS9.7AI score0.02036EPSS
SaveExploits0References1
Prion
Prion
•added 2019/03/25 4:29 p.m.•29 views

Authorization

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

5CVSS6.6AI score0.01008EPSS
SaveExploits0References1Affected Software3
NVD
NVD
•added 2019/03/25 4:29 p.m.•38 views

CVE-2015-3952

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

7.5CVSS8AI score0.01008EPSS
SaveExploits0References1
Cvelist
Cvelist
•added 2019/03/25 4:02 p.m.•35 views

CVE-2015-3953

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

9.5AI score0.01972EPSS
SaveExploits0References1
Rapid7 Vulnerability Database (full)
Rapid7 Vulnerability Database (full)
•added 2019/03/25 12:00 a.m.•5 views

CVE-2015-3954: Improper Authorization

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira...

10CVSS5.8AI score0.02036EPSS
SaveExploits0References3
Rapid7 Vulnerability Database (full)
Rapid7 Vulnerability Database (full)
•added 2019/03/25 12:00 a.m.•7 views

CVE-2015-3953: Use of Hard-coded Password

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

10CVSS5.7AI score0.01972EPSS
SaveExploits0References3
Positive Technologies
Positive Technologies
•added 2019/03/25 12:00 a.m.•5 views

PT-2019-7183

Name of the Vulnerable Software and Affected Versions Hospira Plum A+ Infusion System versions 13.4 and prior Hospira Plum A+3 Infusion System versions 13.6 and prior Hospira Symbiq Infusion System version 3.13 and prior Description The issue allows unauthenticated users to gain root privileges o...

10CVSS0.02036EPSS
SaveExploits0References2
Positive Technologies
Positive Technologies
•added 2019/03/25 12:00 a.m.•6 views

PT-2019-7182

Name of the Vulnerable Software and Affected Versions Hospira Plum A+ Infusion System versions 13.4 and prior Hospira Plum A+3 Infusion System versions 13.6 and prior Hospira Symbiq Infusion System version 3.13 and prior Description The issue allows access to the system through hard-coded account...

10CVSS0.01972EPSS
SaveExploits0References2
Rows per page
Query Builder