Hardcoded credentials

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

10CVSS6.9AI score0.01972EPSS
SaveExploits0References1Affected Software3
Prion
Prion
•added 2019/03/25 4:29 p.m.•29 views

Authorization

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

5CVSS6.6AI score0.01008EPSS
SaveExploits0References1Affected Software3
NVD
NVD
•added 2019/03/25 4:29 p.m.•39 views

CVE-2015-3952

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

7.5CVSS8AI score0.01008EPSS
SaveExploits0References1
Cvelist
Cvelist
•added 2019/03/25 4:02 p.m.•37 views

CVE-2015-3953

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

9.5AI score0.01972EPSS
SaveExploits0References1
Rapid7 Vulnerability Database (full)
Rapid7 Vulnerability Database (full)
•added 2019/03/25 12:00 a.m.•7 views

CVE-2015-3953: Use of Hard-coded Password

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...

10CVSS5.7AI score0.01972EPSS
SaveExploits0References3
Positive Technologies
Positive Technologies
•added 2019/03/25 12:00 a.m.•6 views

PT-2019-7182

Name of the Vulnerable Software and Affected Versions Hospira Plum A+ Infusion System versions 13.4 and prior Hospira Plum A+3 Infusion System versions 13.6 and prior Hospira Symbiq Infusion System version 3.13 and prior Description The issue allows access to the system through hard-coded account...