11 matches found
EUVD-2015-3985
Malware in sbrugna...
EUVD-2015-3984
Malware in sbrugna...
Design/Logic Flaw
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host...
Design/Logic Flaw
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira...
Hardcoded credentials
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...
CVE-2015-3953
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...
Authorization
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...
CVE-2015-3952
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...
CVE-2015-3953
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices...
CVE-2015-3953
CVE-2015-3953 affects Hospira Plum A+ Infusion System (versions prior to 13.4), Plum A+3 Infusion System (prior to 13.6), and Symbiq Infusion System (prior to 3.13). Root cause: hard-coded credentials allowing access via affected devices; additional risk factors include services listening on Port...
Hospira Plum A+ and Symbiq Infusion Systems Vulnerabilities
OVERVIEW Independent researcher Billy Rios has identified vulnerabilities in Hospira’s Plum A+ Infusion System that are similar to vulnerabilities identified in Hospira’s LifeCare PCA Infusion System discussed in advisory, ICSA-15-125-01B Hospira LifeCare PCA Infusion System Vulnerabilities...