CVE-2024-48967 Life2000 ventilator and Service PC lack sufficient audit logging capabilities

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings...

10CVSS0.00612EPSS
Exploits0References1
Vulnrichment
Vulnrichment
added 2024/11/14 9:31 p.m.14 views

CVE-2024-48970 Life2000 Ventilator microcontroller lacks memory protection

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure...

9.3CVSS6.5AI score0.00214EPSS
Exploits0References1
Cvelist
Cvelist
added 2024/11/14 9:31 p.m.18 views

CVE-2024-48970 Life2000 Ventilator microcontroller lacks memory protection

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure...

9.3CVSS0.00214EPSS
Exploits0References1
Cvelist
Cvelist
added 2024/11/14 9:27 p.m.20 views

CVE-2024-48974 Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This...

9.3CVSS0.00155EPSS
Exploits0References1
Vulnrichment
Vulnrichment
added 2024/11/14 9:27 p.m.15 views

CVE-2024-48974 Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This...

9.3CVSS6.4AI score0.00155EPSS
Exploits0References1
Vulnrichment
Vulnrichment
added 2024/11/14 9:24 p.m.12 views

CVE-2024-48973 Debug port on Life2000 Ventilator serial interface is enabled by default

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port which are unencrypted; see 3.2.1 that result in unauthorized disclosure of information and/or have unintended impacts on device settings and...

9.3CVSS6.4AI score0.00221EPSS
Exploits0References1
Cvelist
Cvelist
added 2024/11/14 9:24 p.m.18 views

CVE-2024-48973 Debug port on Life2000 Ventilator serial interface is enabled by default

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port which are unencrypted; see 3.2.1 that result in unauthorized disclosure of information and/or have unintended impacts on device settings and...

9.3CVSS0.00221EPSS
Exploits0References1
Cvelist
Cvelist
added 2024/11/14 9:13 p.m.18 views

CVE-2024-48971 Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator

The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges...

9.3CVSS0.00217EPSS
Exploits0References1
Vulnrichment
Vulnrichment
added 2024/11/14 9:13 p.m.13 views

CVE-2024-48971 Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator

The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges...

9.3CVSS7.2AI score0.00217EPSS
Exploits0References1
CNNVD
CNNVD
added 2024/11/14 12:0 a.m.3 views

Baxter Life2000 安全漏洞

Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. A security vulnerability exists in Baxter Life2000 version 06.08.00.00 and prior versions, which originated when enabled by default, that allows sending and receiving of unencrypted messages, which could result in unauthorized...