20 matches found
CVE-2024-48967 Life2000 ventilator and Service PC lack sufficient audit logging capabilities
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings...
CVE-2024-48966 Life2000 service tools for test and calibration do not support user authentication
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedde...
CVE-2024-48966 Life2000 service tools for test and calibration do not support user authentication
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedde...
CVE-2024-48970 Life2000 Ventilator microcontroller lacks memory protection
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure...
CVE-2024-48970 Life2000 Ventilator microcontroller lacks memory protection
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure...
CVE-2024-48974 Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This...
CVE-2024-48974 Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This...
CVE-2024-48973 Debug port on Life2000 Ventilator serial interface is enabled by default
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port which are unencrypted; see 3.2.1 that result in unauthorized disclosure of information and/or have unintended impacts on device settings and...
CVE-2024-48973 Debug port on Life2000 Ventilator serial interface is enabled by default
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port which are unencrypted; see 3.2.1 that result in unauthorized disclosure of information and/or have unintended impacts on device settings and...
CVE-2024-48971 Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges...
CVE-2024-48971 Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges...
CVE-2024-9834 Improper data protection on Life2000 ventilator serial interface
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance...
Baxter Life2000 安全漏洞
The Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. A security vulnerability exists in Baxter Life2000 version 06.08.00.00 and prior versions, which stems from the lack of sufficient audit logging functionality to detect malicious activity and perform subsequent forensic...
Baxter Life2000 安全漏洞
The Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. A security vulnerability exists in Baxter Life2000 version 06.08.00.00 and earlier, which stems from a lack of memory protection and allows an attacker to read or write to flash memory via the JTAG interface, potentially...
Baxter Life2000 安全漏洞
Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. A security vulnerability exists in Baxter Life2000 version 06.08.00.00 and prior versions, which originated when enabled by default, that allows sending and receiving of unencrypted messages, which could result in unauthorized...
Baxter Life2000 安全漏洞
The Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. A security vulnerability exists in Baxter Life2000 version 06.08.00.00 and earlier, which stems from improper file integrity checking when applying a firmware update, allowing an attacker to push a compromised or illegal...
Baxter Life2000 安全漏洞
The Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. A security vulnerability exists in the Baxter Life2000 that stems from improper data protection on the ventilator's serial interface, which could allow an attacker to send and receive messages that could have an unintended...
Baxter Life2000 访问控制错误漏洞
The Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. An access control error vulnerability exists in Baxter Life2000 version 06.08.00.00 and prior, which stems from unsupported user authentication, allowing an attacker to obtain diagnostic information or manipulate device...
Baxter Life2000 信任管理问题漏洞
The Baxter Life2000 is a mask-less non-invasive ventilator from Baxter. A trust management issue vulnerability exists in Baxter Life2000 version 06.08.00.00 and prior versions, which stems from the clinician password and serial number clinician password, hardcoded in plaintext in the device, whic...
Baxter Life2000 安全漏洞
The Baxter Life2000 is a mask-less noninvasive ventilator from Baxter. A security vulnerability exists in the Baxter Life2000 that stems from an unlimited number of login failures allowed using the clinician password or serial number clinician password, which allows an attacker to perform a brute...