183 matches found
Default configuration
Fresenius Kabi Agilia SP MC WiFi vD25 and prior has a default configuration page accessible without authentication. An attacker may use this functionality to change the exposed configuration values such as network settings...
Information disclosure
The SSL/TLS configuration of Fresenius Kabi Agilia Link + version 3.0 has serious deficiencies that may allow an attacker to compromise SSL/TLS sessions in different ways. An attacker may be able to eavesdrop on transferred data, manipulate data allegedly secured by SSL/TLS, and impersonate an...
Design/Logic Flaw
Fresenius Kabi Vigilant Software Suite Mastermed Dashboard version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypa...
Design/Logic Flaw
An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating user...
CVE-2021-41835 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect from the unencrypted service on Port...
CVE-2021-43355 Fresenius Kabi Agilia Connect Infusion System use of client side authentication
Fresenius Kabi Vigilant Software Suite Mastermed Dashboard version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypa...
CVE-2021-43355 Fresenius Kabi Agilia Connect Infusion System use of client side authentication
Fresenius Kabi Vigilant Software Suite Mastermed Dashboard version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypa...
CVE-2021-43355
The CVE-2021-43355 entry concerns Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) v2.0.1.3, where client-side input validation is performed without server-side authentication, allowing a user to bypass checks and login with service privileges. The root cause is misuse of client-side ...
CVE-2021-41835
CVE-2021-41835 affects Fresenius Kabi Agilia Link+ and related components in Agilia Connect Infusion System. The root issue is failure to enforce transport-layer encryption: data may traverse plaintext, with TLS offered on port 443 but no automatic redirect from unencrypted port 80 to 443. The vu...
CVE-2021-41835 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect from the unencrypted service on Port...
CVE-2021-33848 Fresenius Kabi Agilia Connect Infusion System cross site scripting
Fresenius Kabi Vigilant Software Suite Mastermed Dashboard version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing internal information and performing actions...
CVE-2021-33848
CVE-2021-33848 affects Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3, with a reflected cross-site scripting vulnerability described for the Vigilant Centerium Dashboard. An attacker could inject JavaScript via a GET parameter in HTTP requests, enabling unauthorized ...
CVE-2021-33848 Fresenius Kabi Agilia Connect Infusion System cross site scripting
Fresenius Kabi Vigilant Software Suite Mastermed Dashboard version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing internal information and performing actions...
CVE-2021-44464 Fresenius Kabi Agilia Connect Infusion System hard coded credentials
Vigilant Software Suite Mastermed Dashboard version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software...
CVE-2021-44464 Fresenius Kabi Agilia Connect Infusion System hard coded credentials
Vigilant Software Suite Mastermed Dashboard version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software...
CVE-2021-33843 Fresenius Kabi Agilia Connect Infusion System files or directories accessible to external parties
Fresenius Kabi Agilia SP MC WiFi vD25 and prior has a default configuration page accessible without authentication. An attacker may use this functionality to change the exposed configuration values such as network settings...
CVE-2021-33843 Fresenius Kabi Agilia Connect Infusion System files or directories accessible to external parties
Fresenius Kabi Agilia SP MC WiFi vD25 and prior has a default configuration page accessible without authentication. An attacker may use this functionality to change the exposed configuration values such as network settings...
CVE-2021-33843
CVE-2021-33843 concerns Fresenius Kabi Agilia Connect Infusion System components (Agilia SP MC WiFi vD25 and earlier; Agilia Link+ v3.0 and earlier) with a default configuration page that is accessible without authentication. The vulnerability allows an attacker to alter exposed configuration val...
CVE-2021-31562 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
The SSL/TLS configuration of Fresenius Kabi Agilia Link + version 3.0 has serious deficiencies that may allow an attacker to compromise SSL/TLS sessions in different ways. An attacker may be able to eavesdrop on transferred data, manipulate data allegedly secured by SSL/TLS, and impersonate an...
CVE-2021-31562 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
The SSL/TLS configuration of Fresenius Kabi Agilia Link + version 3.0 has serious deficiencies that may allow an attacker to compromise SSL/TLS sessions in different ways. An attacker may be able to eavesdrop on transferred data, manipulate data allegedly secured by SSL/TLS, and impersonate an...