CVE-2021-33017
The standard access path of the IntelliBridge EC 40 and 60 Hub C.00.04 and prior requires authentication, but the product has an alternate path or channel that does not require authentication...
CVE-2021-32993
IntelliBridge EC 40 and 60 Hub C.00.04 and prior contains hard-coded credentials, such as a password or a cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data...
Authentication flaw
The standard access path of the IntelliBridge EC 40 and 60 Hub C.00.04 and prior requires authentication, but the product has an alternate path or channel that does not require authentication...
CVE-2021-33017 Philips IntelliBridge EC 40 and EC 80 Hub Authentication Bypass Using an Alternate Path or Channel
The standard access path of the IntelliBridge EC 40 and 60 Hub C.00.04 and prior requires authentication, but the product has an alternate path or channel that does not require authentication...
CVE-2021-33017
CVE-2021-33017 affects Philips IntelliBridge EC40 Hub (C.00.04 and prior) and EC80 Hub, exposing an authentication bypass via an alternate path or channel that does not require authentication. The ICS-CISA advisory documents an authentication bypass vulnerability (CWE-288) with CVSS v3 base score...
CVE-2021-32993
The CVE-2021-32993 issue affects Philips IntelliBridge EC 40 Hub (C.00.04 and prior) and EC 80 Hub (C.00.04 and prior). The vulnerability arises from hard-coded credentials used for inbound authentication, outbound communication to external components, or encryption of internal data (CWE-798). Re...
Philips IntelliBridge EC 40 and EC 80 Hub
1. EXECUTIVE SUMMARY CVSS v3 8.1 ATTENTION: Low attack complexity Vendor: Philips Equipment: IntelliBridge EC 40 and EC 80 Hub Vulnerabilities: Use of Hard-coded Credentials, Authentication Bypass Using an Alternate Path or Channel 2. RISK EVALUATION Successful exploitation of these...