2 matches found
CVE-2020-12035
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings...
4.9CVSS7AI score0.00247EPSS
SaveExploits0
CVE-2020-12035
CVE-2020-12035 affects Baxter PrismaFlex (all versions) and PrisMax (all versions prior to 3.x). Root cause is a hard-coded service password that grants access to biomedical information, device settings, calibration settings, and network configuration, enabling an attacker to modify device settin...
4.9CVSS5.1AI score0.00247EPSS
SaveExploits0References2Affected Software1