In GE Aestiva and Aespire versions 7100 and 7900, a vulnerability exists where serial devices are connected via an added unsecured terminal server to a TCP/IP network configuration, which could allow an attacker to remotely modify device configuration and silence alarms...
5.3CVSS6.9AI score0.01336EPSS
SaveExploits0References1
CVE-2019-10966
CVE-2019-10966 affects GE Aestiva and Aespire anesthesia devices (7100/7900) where serial ports extended to TCP/IP via an unsecured terminal server. Root cause is improper authentication, enabling remote modification of device configuration and silencing of alarms without user interaction. The IC...
5.3CVSS5.2AI score0.01336EPSS
SaveExploits0References2Affected Software1