CVE-2017-12722
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the...
CVE-2017-12722
CVE-2017-12722 affects Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump (versions 1.1, 1.5, 1.6). A DNS client in the MQX RTOS reads memory out-of-bounds, potentially crashing the communications module (remote exposure). Public docs consistently state this may be exploited remotely an...
CVE-2017-12722
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the...
NXP Semiconductors MQX RTOS contains multiple vulnerabilities
Overview The NXP Semiconductors MQX RTOS prior to version 5.1 contains a buffer overflow in the DHCP client, which may lead to memory corruption allowing an attacker to execute arbitrary code, as well as an out of bounds read in the DNS client which may lead to a denial of service. Description Th...
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A)
OVERVIEW This updated advisory is a follow-up to the original advisory titled ICSMA-17-250-02 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities that was published September 7, 2017, on the NCCIC/ICS-CERT web site. Independent researcher Scott Gayou has identified eight...