Lucene search
K

4 matches found

OSV
OSV
added 2018/04/25 1:29 p.m.4 views

CVE-2017-12716

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CV...

6.5CVSS5.8AI score0.00248EPSS
Exploits0References2
OSV
OSV
added 2018/04/25 1:29 p.m.3 views

CVE-2017-12712

The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3...

8.8CVSS5.8AI score0.01084EPSS
Exploits0References2
CNVD
CNVD
added 2017/08/30 12:0 a.m.5 views

Unauthorized Access Vulnerability in Various Abbott Laboratories Pacemaker Products

Accent, Anthem, Accent MRI, Assurity, Allure and Assurity MRI are implantable medical devices from Abbott Laboratories USA. An unauthorized access vulnerability exists in several Abbott Laboratories pacemaker products, where pacemaker authentication algorithms involving authentication keys and...

8.8CVSS8.9AI score0.01084EPSS
Exploits0References1
CNVD
CNVD
added 2017/08/30 12:0 a.m.3 views

Abbott Laboratories Multiple Pacemaker Products Access Count Limit Vulnerability

Accent, Anthem, Accent MRI, Assurity, Allure and Assurity MRI are implantable medical devices from Abbott Laboratories USA. The access limit vulnerability exists in several Abbott Laboratories pacemaker products, where the pacemaker does not limit or restricts the number of properly formatted "RF...

6.5CVSS6.9AI score0.00655EPSS
Exploits0References1