•added 2018/07/02 12:00 a.m.•5 views

CVE-2018-8870: Use of Hard-coded Password

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system...

7.2CVSS6.8AI score0.00362EPSS
SaveExploits0References3
Rapid7 Vulnerability Database (full)
Rapid7 Vulnerability Database (full)
•added 2018/07/02 12:00 a.m.•6 views

CVE-2018-8868: Exposed Dangerous Method or Function

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit oth...

6.9CVSS6.8AI score0.00333EPSS
SaveExploits0References2