42 matches found
EUVD-2021-23738
Malware in sbrugna...
EUVD-2021-23734
Malware in sbrugna...
EUVD-2021-23736
Malware in sbrugna...
EUVD-2021-23737
Malware in sbrugna...
EUVD-2021-23735
Malware in sbrugna...
CVE-2021-37167
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. A user logged in using the default credentials can gain root access to the device, which provides permissions for all of t...
CVE-2021-37163
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus operated by released versions of software before Nexus Software 7.2.5.7. The device has two user accounts with passwords that are hardcoded...
CVE-2021-37164
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a...
CVE-2021-37160
A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation e.g., cryptographic signature validation during a File Upload for a firmware update...
CVE-2021-37166
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and...
CVE-2021-37162
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote co...
HMI3 Control Panel Trust Management Issue Vulnerability
Swisslog Healthcare Nexus Panel is a medical device from Swisslog Healthcare.A security vulnerability exists in the HMI3 Control Panel, which stems from an insecure privilege issue in the HMI3 Control Panel of the Swisslog Healthcare Nexus. No details of the vulnerability are currently available...
CVE-2021-37164
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a...
CVE-2021-37164
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a...
CVE-2021-37160
A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation e.g., cryptographic signature validation during a File Upload for a firmware update...
CVE-2021-37160
A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation e.g., cryptographic signature validation during a File Upload for a firmware update...
CVE-2021-37163
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus operated by released versions of software before Nexus Software 7.2.5.7. The device has two user accounts with passwords that are hardcoded...
CVE-2021-37161
A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel, operated by released versions of software before Nexus Software 7.2.5.7. A buffer overflow allows an attacker to overwrite an internal queue data structure and can lead to remote...
CVE-2021-37162
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote co...
CVE-2021-37161
A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel, operated by released versions of software before Nexus Software 7.2.5.7. A buffer overflow allows an attacker to overwrite an internal queue data structure and can lead to remote...