2 matches found
CVE-2026-18164
The CVE concerns Flow Neuroscience FL-100 with an undocumented hard-coded credential that bypasses authentication. The credential is shared across all device units, allowing an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state. Affected component is ...
8.1CVSS5.8AI score0.00388EPSS
SaveExploits0References2
Flow Neuroscience FL-100
ADVISORY SUMMARY Successful exploitation of this vulnerability could allow an attacker within Bluetooth range to manipulate brain stimulation parameters and override safety limits. 2. RECOMMENDED PRACTICES CISA recommends users take defensive measures to minimize the risk of exploitation of this...
8.1CVSS5.8AI score0.00388EPSS
SaveExploits0References13