4 matches found
CVE-2020-12036
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption e.g., TLS/SSL when configured to send treatment data to a PDMS Patient Data Management System or an EMR Electronic Medical Record system. An attacker could observe...
CVE-2020-12036
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption e.g., TLS/SSL when configured to send treatment data to a PDMS Patient Data Management System or an EMR Electronic Medical Record system. An attacker could observe...
CVE-2020-12036
CVE-2020-12036 affects Baxter PrismaFlex (all versions) and PrisMax (all versions prior to 3.x). The root cause is lack of data-in-transit encryption (no TLS/SSL) when these devices send treatment data to a PDMS or EMR, enabling an attacker with network access to observe sensitive data. The ICS a...
CVE-2020-12036
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption e.g., TLS/SSL when configured to send treatment data to a PDMS Patient Data Management System or an EMR Electronic Medical Record system. An attacker could observe...