12 matches found
CVE-2019-18246
BIOTRONIK CardioMessenger II, The affected products do not properly enforce mutual authentication with the BIOTRONIK Remote Communication infrastructure...
CVE-2019-18252
BIOTRONIK CardioMessenger II, The affected products allow credential reuse for multiple authentication purposes. An attacker with adjacent access to the CardioMessenger can disclose its credentials used for connecting to the BIOTRONIK Remote Communication infrastructure...
CVE-2019-18254
BIOTRONIK CardioMessenger II, The affected products do not encrypt sensitive information while at rest. An attacker with physical access to the CardioMessenger can disclose medical measurement data and the serial number from the implanted cardiac device the CardioMessenger is paired with...
CVE-2019-18248
BIOTRONIK CardioMessenger II, The affected products transmit credentials in clear-text prior to switching to an encrypted communication channel. An attacker can disclose the product’s client credentials for connecting to the BIOTRONIK Remote Communication infrastructure...
CVE-2019-18252
BIOTRONIK CardioMessenger II, The affected products allow credential reuse for multiple authentication purposes. An attacker with adjacent access to the CardioMessenger can disclose its credentials used for connecting to the BIOTRONIK Remote Communication infrastructure...
CVE-2019-18256
BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit...
CVE-2019-18256: Storing Passwords in a Recoverable Format
BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit...
CVE-2019-18252: Improper Authentication
BIOTRONIK CardioMessenger II, The affected products allow credential reuse for multiple authentication purposes. An attacker with adjacent access to the CardioMessenger can disclose its credentials used for connecting to the BIOTRONIK Remote Communication infrastructure...
CVE-2019-18254: Missing Encryption of Sensitive Data
BIOTRONIK CardioMessenger II, The affected products do not encrypt sensitive information while at rest. An attacker with physical access to the CardioMessenger can disclose medical measurement data and the serial number from the implanted cardiac device the CardioMessenger is paired with...
CVE-2019-18246: Improper Authentication
BIOTRONIK CardioMessenger II, The affected products do not properly enforce mutual authentication with the BIOTRONIK Remote Communication infrastructure...
CVE-2019-18248: Cleartext Transmission of Sensitive Information
BIOTRONIK CardioMessenger II, The affected products transmit credentials in clear-text prior to switching to an encrypted communication channel. An attacker can disclose the product’s client credentials for connecting to the BIOTRONIK Remote Communication infrastructure...
BIOTRONIK CardioMessenger II
1. EXECUTIVE SUMMARY CVSS v3 4.6 ATTENTION: Exploitable with adjacent access/low skill level to exploit Vendor: BIOTRONIK Equipment: CardioMessenger II-S T-Line, CardioMessenger II-S GSM Vulnerabilities: Improper Authentication, Cleartext Transmission of Sensitive Information, Missing Encryption...