Lucene search

K

Mycarelink 24950 Patient Monitor Firmware Security Vulnerabilities

cve
cve

CVE-2018-10622

A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected products use per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication and encryption of local data at rest.

7.1CVSS

6.9AI Score

0.001EPSS

2018-08-10 06:29 PM
26
cve
cve

CVE-2018-10626

A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product's update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable c...

4.4CVSS

4.7AI Score

0.0004EPSS

2018-08-10 06:29 PM
30