Lucene search

K
icsIndustrial Control Systems Cyber Emergency Response TeamICSMA-23-285-02
HistoryOct 12, 2023 - 12:00 p.m.

Santesoft Sante FFT Imaging

2023-10-1212:00:00
Industrial Control Systems Cyber Emergency Response Team
www.cisa.gov
20
santesoft
fft imaging
vulnerability
out-of-bounds read
exploitation
cvss v3 7.8
dicom
healthcare
public health
cyprus
cwe-125
cve-2023-5059
cisa
mitigations
control systems
cybersecurity
defensive measures
vpns
remote access
impact analysis
risk assessment

7.8 High

CVSS3

Attack Vector

LOCAL

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

REQUIRED

Scope

UNCHANGED

Confidentiality Impact

HIGH

Integrity Impact

HIGH

Availability Impact

HIGH

CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H

0.001 Low

EPSS

Percentile

32.3%

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v3 7.8 *ATTENTION: Low attack complexity
  • Vendor: Santesoft
  • Equipment: Sante FFT Imaging
  • Vulnerability: Out-of-Bounds Read

2. RISK EVALUATION

Successful exploitation of this vulnerability could allow an attacker to disclose information and execute arbitrary code.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Santesof products are affected:

  • Sante FFT Imaging: Versions v1.4.0 and prior

3.2 Vulnerability Overview

3.2.1 OUT-OF-BOUNDS READ CWE-125

The affected application lacks proper validation of user-supplied data when parsing DICOM files. This could lead to an out-of-bounds read. An attacker could leverage this vulnerability to execute arbitrary code in the context of the current process.

CVE-2023-5059 has been assigned to this vulnerability. A CVSS v3 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Cyprus

3.4 RESEARCHER

Michael Heinzl reported this vulnerability to CISA.

4. MITIGATIONS

Santesoft has released an updated version of their product and recommends users update to Sante FFT Imaging to v1.4.1.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

5. UPDATE HISTORY

  • October 12, 2023: Initial Publication

References

7.8 High

CVSS3

Attack Vector

LOCAL

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

REQUIRED

Scope

UNCHANGED

Confidentiality Impact

HIGH

Integrity Impact

HIGH

Availability Impact

HIGH

CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H

0.001 Low

EPSS

Percentile

32.3%

Related for ICSMA-23-285-02