CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
HIGH
User Interaction
NONE
Scope
UNCHANGED
Confidentiality Impact
NONE
Integrity Impact
HIGH
Availability Impact
NONE
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
AI Score
Confidence
High
EPSS
Percentile
35.4%
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
[
{
"defaultStatus": "unaffected",
"product": "Space Battery Pack SP with Wi-Fi",
"vendor": "B. Braun ",
"versions": [
{
"lessThanOrEqual": "053L000092",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "054U000092",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
]
CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
HIGH
User Interaction
NONE
Scope
UNCHANGED
Confidentiality Impact
NONE
Integrity Impact
HIGH
Availability Impact
NONE
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
AI Score
Confidence
High
EPSS
Percentile
35.4%