Lucene search

K
cvelistIcscertCVELIST:CVE-2022-38453
HistorySep 13, 2022 - 2:55 p.m.

CVE-2022-38453 Contec Health CMS8000

2022-09-1314:55:03
CWE-489
icscert
www.cve.org
2
contec health cms8000
'not stripped'
'debug_info'
vulnerability
reverse engineer
sensitive code
additional vulnerabilities

CVSS3

3

Attack Vector

LOCAL

Attack Complexity

HIGH

Privileges Required

HIGH

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

LOW

Integrity Impact

LOW

Availability Impact

NONE

CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N

AI Score

5.2

Confidence

High

EPSS

0

Percentile

10.4%

Multiple binary application files on the CMS8000 device are compiled with ‘not stripped’ and ‘debug_info’ compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.

CNA Affected

[
  {
    "product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
    "vendor": "Contec Health",
    "versions": [
      {
        "status": "affected",
        "version": "All"
      }
    ]
  }
]

CVSS3

3

Attack Vector

LOCAL

Attack Complexity

HIGH

Privileges Required

HIGH

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

LOW

Integrity Impact

LOW

Availability Impact

NONE

CVSS:3.1/AV:L/AC:H/PR:H/UI:N/S:U/C:L/I:L/A:N

AI Score

5.2

Confidence

High

EPSS

0

Percentile

10.4%

Related for CVELIST:CVE-2022-38453