Lucene search

K
cvelistSiemensCVELIST:CVE-2022-29875
HistoryJun 01, 2022 - 9:50 a.m.

CVE-2022-29875

2022-06-0109:50:11
CWE-502
siemens
www.cve.org
3
biograph horizon pet/ct
magnetom family
mammomat revelation
naeotom alpha
somatom x.cite
somatom x.creed
somatom go.all
somatom go.now
somatom go.open pro
somatom go.sim
somatom go.top
somatom go.up
symbia e/s
symbia evo
symbia intevo
symbia t
symbia.net
syngo.via vb10
syngo.via vb20
syngo.via vb30
syngo.via vb40
syngo.via vb50
syngo.via vb60
arbitrary deserialization
unauthenticated attacker
code execution
32912/tcp
32914/tcp

AI Score

9.7

Confidence

High

EPSS

0.002

Percentile

59.2%

A vulnerability has been identified in Biograph Horizon PET/CT Systems (All VJ30 versions < VJ30C-UD01), MAGNETOM Family (NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A), MAMMOMAT Revelation (All VC20 versions < VC20D), NAEOTOM Alpha (All VA40 versions < VA40 SP2), SOMATOM X.cite (All versions < VA30 SP5 or VA40 SP2), SOMATOM X.creed (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.All (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Now (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Open Pro (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Sim (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Top (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Up (All versions < VA30 SP5 or VA40 SP2), Symbia E/S (All VB22 versions < VB22A-UD03), Symbia Evo (All VB22 versions < VB22A-UD03), Symbia Intevo (All VB22 versions < VB22A-UD03), Symbia T (All VB22 versions < VB22A-UD03), Symbia.net (All VB22 versions < VB22A-UD03), syngo.via VB10 (All versions), syngo.via VB20 (All versions), syngo.via VB30 (All versions), syngo.via VB40 (All versions < VB40B HF06), syngo.via VB50 (All versions), syngo.via VB60 (All versions < VB60B HF02). The application deserialises untrusted data without sufficient validations that could result in an arbitrary deserialization. This could allow an unauthenticated attacker to execute code in the affected system if ports 32912/tcp or 32914/tcp are reachable.

CNA Affected

[
  {
    "product": "Biograph Horizon PET/CT Systems",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VJ30 versions < VJ30C-UD01"
      }
    ]
  },
  {
    "product": "MAGNETOM Family",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A"
      }
    ]
  },
  {
    "product": "MAMMOMAT Revelation",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VC20 versions < VC20D"
      }
    ]
  },
  {
    "product": "NAEOTOM Alpha",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VA40 versions < VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM X.cite",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM X.creed",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM go.All",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM go.Now",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM go.Open Pro",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM go.Sim",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM go.Top",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "SOMATOM go.Up",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VA30 SP5 or VA40 SP2"
      }
    ]
  },
  {
    "product": "Symbia E/S",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VB22 versions < VB22A-UD03"
      }
    ]
  },
  {
    "product": "Symbia Evo",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VB22 versions < VB22A-UD03"
      }
    ]
  },
  {
    "product": "Symbia Intevo",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VB22 versions < VB22A-UD03"
      }
    ]
  },
  {
    "product": "Symbia T",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VB22 versions < VB22A-UD03"
      }
    ]
  },
  {
    "product": "Symbia.net",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All VB22 versions < VB22A-UD03"
      }
    ]
  },
  {
    "product": "syngo.via VB10",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions"
      }
    ]
  },
  {
    "product": "syngo.via VB20",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions"
      }
    ]
  },
  {
    "product": "syngo.via VB30",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions"
      }
    ]
  },
  {
    "product": "syngo.via VB40",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VB40B HF06"
      }
    ]
  },
  {
    "product": "syngo.via VB50",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions"
      }
    ]
  },
  {
    "product": "syngo.via VB60",
    "vendor": "Siemens",
    "versions": [
      {
        "status": "affected",
        "version": "All versions < VB60B HF02"
      }
    ]
  }
]

AI Score

9.7

Confidence

High

EPSS

0.002

Percentile

59.2%

Related for CVELIST:CVE-2022-29875