Lucene search

K
cvelistIcscertCVELIST:CVE-2021-38392
HistorySep 30, 2021 - 12:00 a.m.

CVE-2021-38392 Improper Access Control for Boston Scientific Zoom Latitude

2021-09-3000:00:00
CWE-284
icscert
www.cve.org

6.5 Medium

CVSS3

Attack Vector

PHYSICAL

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

REQUIRED

Scope

CHANGED

Confidentiality Impact

LOW

Integrity Impact

HIGH

Availability Impact

LOW

CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L

7.6 High

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

27.5%

A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.

CNA Affected

[
  {
    "product": "ZOOM LATITUDE",
    "vendor": "Boston Scientific",
    "versions": [
      {
        "status": "affected",
        "version": "Model 3120"
      }
    ]
  }
]

6.5 Medium

CVSS3

Attack Vector

PHYSICAL

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

REQUIRED

Scope

CHANGED

Confidentiality Impact

LOW

Integrity Impact

HIGH

Availability Impact

LOW

CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L

7.6 High

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

27.5%

Related for CVELIST:CVE-2021-38392