HistoryDec 14, 2020 - 7:18 p.m.

CVE-2020-25183

2020-12-1419:18:44
CWE-287
icscert
www.cve.org

8.8 High

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

28.7%

Medtronic MyCareLink Smart 25000 all versions contain an authentication protocol vuln where the method used to auth between MCL Smart Patient Reader and MyCareLink Smart mobile app is vulnerable to bypass. This vuln allows attacker to use other mobile device or malicious app on smartphone to auth to the patient’s Smart Reader, fools the device into thinking its communicating with the actual smart phone application when executed in range of Bluetooth.

CNA Affected

[
  {
    "product": "Medtronic MyCareLink Smart 25000 Reader",
    "vendor": "n/a",
    "versions": [
      {
        "status": "affected",
        "version": "Smart 25000 all versions"
      }
    ]
  }
]