cvelistIcscertCVELIST:CVE-2018-14781
HistoryAug 13, 2018 - 10:00 p.m.

CVE-2018-14781

2018-08-1322:00:00
CWE-294
icscert
www.cve.org
3

AI Score

5.3

Confidence

High

EPSS

0.001

Percentile

30.5%

Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

CNA Affected

[
  {
    "product": "Medtronic insulin pump",
    "vendor": "ICS-CERT",
    "versions": [
      {
        "status": "affected",
        "version": "MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G"
      }
    ]
  }
]