CVE-2017-12712

2017-08-2900:00:00
CWE-287
icscert
www.cve.org
3

8.8 High

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

32.2%

The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CNA Affected

[
  {
    "product": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
    "vendor": "Abbott Laboratories",
    "versions": [
      {
        "status": "affected",
        "version": "All versions of pacemakers manufactured prior to August 28, 2017"
      }
    ]
  }
]