Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CNA Affected

[
  {
    "product": "Plum A+ Infusion System",
    "vendor": "Hospira",
    "versions": [
      {
        "status": "affected",
        "version": "<= 13.4"
      }
    ]
  },
  {
    "product": "Plum A+3 Infusion System",
    "vendor": "Hospira",
    "versions": [
      {
        "status": "affected",
        "version": "<= 13.6"
      }
    ]
  },
  {
    "product": "Symbiq Infusion System",
    "vendor": "Hospira",
    "versions": [
      {
        "status": "affected",
        "version": "<= 3.13"
      }
    ]
  }
]