Lucene search
K

CVE-2024-48971

🗓️ 14 Nov 2024 21:13:36Reported by BaxterType 
cve
 cve
🔗 web.nvd.nist.gov👁 44 Views

The ventilator has hard-coded clinician passwords and serial numbers in plaintext, allowing unauthorized access

Related
Affected
Refs
[
  {
    "defaultStatus": "unaffected",
    "product": "Life2000 Ventilation System",
    "vendor": "Baxter",
    "versions": [
      {
        "status": "affected",
        "version": "06.08.00.00 and prior"
      }
    ]
  }
]