Lucene search

K
cve[email protected]CVE-2020-16212
HistorySep 11, 2020 - 2:15 p.m.

CVE-2020-16212

2020-09-1114:15:11
CWE-668
web.nvd.nist.gov
28
cve-2020-16212
patient information center ix
performancebridge focal point
intellivue patient monitors
vulnerability
access control
security breach

4.6 Medium

CVSS2

Attack Vector

LOCAL

Attack Complexity

LOW

Authentication

NONE

Confidentiality Impact

PARTIAL

Integrity Impact

PARTIAL

Availability Impact

PARTIAL

AV:L/AC:L/Au:N/C:P/I:P/A:P

6.8 Medium

CVSS3

Attack Vector

PHYSICAL

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

HIGH

Integrity Impact

HIGH

Availability Impact

HIGH

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

6.3 Medium

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

32.3%

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource. The application on the surveillance station operates in kiosk mode, which is vulnerable to local breakouts that could allow an attacker with physical access to escape the restricted environment with limited privileges.

Affected configurations

NVD
Node
philipspatient_information_center_ixMatchb.02
OR
philipspatient_information_center_ixMatchc.02
OR
philipspatient_information_center_ixMatchc.03

CNA Affected

[
  {
    "defaultStatus": "unaffected",
    "product": "Patient Information Center iX (PICiX)",
    "vendor": "Philips",
    "versions": [
      {
        "status": "affected",
        "version": "B.02"
      },
      {
        "status": "affected",
        "version": "C.02"
      },
      {
        "status": "affected",
        "version": "C.03"
      }
    ]
  }
]

4.6 Medium

CVSS2

Attack Vector

LOCAL

Attack Complexity

LOW

Authentication

NONE

Confidentiality Impact

PARTIAL

Integrity Impact

PARTIAL

Availability Impact

PARTIAL

AV:L/AC:L/Au:N/C:P/I:P/A:P

6.8 Medium

CVSS3

Attack Vector

PHYSICAL

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

HIGH

Integrity Impact

HIGH

Availability Impact

HIGH

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

6.3 Medium

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

32.3%

Related for CVE-2020-16212