Lucene search

K
cveIcscertCVE-2017-9657
HistoryApr 30, 2018 - 3:29 p.m.

CVE-2017-9657

2018-04-3015:29:00
CWE-755
CWE-460
icscert
web.nvd.nist.gov
28
cve-2017-9657
philips
intellivue mx40
wlan
network conditions
central monitoring station
alarm events
treatment delay
cvss vector
software update
vulnerability fix
exceptional conditions vulnerability
access point disconnect.

CVSS2

3.3

Attack Vector

ADJACENT_NETWORK

Attack Complexity

LOW

Authentication

NONE

Confidentiality Impact

NONE

Integrity Impact

NONE

Availability Impact

PARTIAL

AV:A/AC:L/Au:N/C:N/I:N/A:P

CVSS3

6.5

Attack Vector

ADJACENT

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

NONE

Integrity Impact

NONE

Availability Impact

HIGH

CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

AI Score

6.2

Confidence

High

EPSS

0.001

Percentile

45.7%

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Affected configurations

Nvd
Node
philipsintellivue_mx40_firmwareRange<b.06.18
AND
philipsintellivue_mx40Match-
VendorProductVersionCPE
philipsintellivue_mx40_firmware*cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:*
philipsintellivue_mx40-cpe:2.3:h:philips:intellivue_mx40:-:*:*:*:*:*:*:*

CNA Affected

[
  {
    "product": "IntelliVue MX40 Patient Worn Monitor",
    "vendor": "Philips",
    "versions": [
      {
        "status": "affected",
        "version": "IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18"
      }
    ]
  }
]

CVSS2

3.3

Attack Vector

ADJACENT_NETWORK

Attack Complexity

LOW

Authentication

NONE

Confidentiality Impact

NONE

Integrity Impact

NONE

Availability Impact

PARTIAL

AV:A/AC:L/Au:N/C:N/I:N/A:P

CVSS3

6.5

Attack Vector

ADJACENT

Attack Complexity

LOW

Privileges Required

NONE

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

NONE

Integrity Impact

NONE

Availability Impact

HIGH

CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

AI Score

6.2

Confidence

High

EPSS

0.001

Percentile

45.7%

Related for CVE-2017-9657